Specifications

Detector

  • Light Source: LED 480nm
  • Detector: Photomultiplier tube with additional digital signal processing
  • Evaporator Temperature Range:

         PL-ELS 2100 Ambient-120C (1C increments)

         PL-ELS 2100 Ice 10 - 80oC (1oC increments)

  • Nebulizer Temperature Range: Ambient-90C (1C increments)
  • Gas Usage: Up to 3.25 L/minute @60 psi @25C (N2 or Air) - may use a nitrogen generator (see datasheet for more information)
  • Integrated automatic, controlled, gas shut-off valve
  • Minimum Operating Pressure: 60 psi (4-6 bar)
  • Maximum Operating Pressure: 100 psi (6.7 bar)
  • Eluent Flow rate range: 0.2-5.0 ml/min
  • Digital Output: 25 bit (10Hz) via serial port
  • Analog Output: 0-1V FSD
  • Communication: Serial I/O (RS232)
  • Contact closure

    Pump stop: 1 contact closure (1 TTL +ve/1 TTL ve)

Control

  • Graphical Vacuum Fluorescent display: Autozero, 4 button keypad
  • Software: PC based control program
  • Stored Methods: 10 predefined methods
  • Operation Modes: Standby, Run

Environment

  • Operation: 10 to 40C  (50 to 104F); relative humidity from 10-80%

    Avoid positioning in direct sunlight

  • Requires Extraction
  • Power Requirements: 90/120V AC or 220/250V AC 50/60 Hz 2A max

Dimensions and Weight

  • Size: 200 mm wide x 450 mm deep x 415 mm high
  • Weight:

        PL-ELS 2100 Ice 13 kg

        PL-ELS 2100 11 kg

 

Comments to: hplc@varianinc.com
Last updated: 09 November 2009